3 Performance specifications
9
whilst it opens. Pressure compensation will have been performed by the time that the door is completely
open, even following pressure differences.
Quantity and quality of the feed water
The quantity and quality of the feed water is automatically checked before every program start.
Overview of sterilization programs (type B)
The results in this table show which inspections were performed on the steam sterilizer. The marked field
shows compliance with all the applicable sections of the standard EN 13060.
Type tests
Universal-
Program
Quick-
Program B
Quick-
Program S
Gentle-
Program
Prion-
Program
Program type in
accordance with
EN 13060
Type B
Type B
Type S
Type B
Type B
Dynamic pressure test of
the sterilization chamber
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
--
--
X
X
X
--
--
X
X
Simple hollow body
(Hollow body B)
X
X
X
X
X
Product with narrow
lumen (Hollow body A)
X
X
--
X
X
X
X
--
X
X
X
--
--
X
X
Drying solid load
X
X
X
X
X
Drying, porous load
X
--
--
X
X
Sterilization temperature
134 °C
134 °C
134 °C
121 °C
134 °C
Sterilization pressure
2.1 bar
2.1 bar
2.1 bar
1.1 bar
2.1 bar
Sterilization time
5:30 min
5:30 min
3:30 min
20:30 min
20:30 min
X = Complies with all applicable sections of the standard EN 13060