13 Malfunctions
75
Incident
Possible causes
What you can do
131
a) The outlet hose is kinked or installed
sagging.
b) The siphon is blocked or the waste water
line is impeded on the building side.
c) Multiple devices have been connected to a
single siphon and imped each other during
water discharge.
d) The chamber filter "Pressure release" is
blocked.
a) Check the installation of the outlet hose.
This must be installed without kinking or
sagging and at a constant decline.
b) Check whether the building siphon or the
waste water line is are blocked.
c) If multiple devices are operated
simultaneously, we recommend the
installation of an additional siphon.
d) Remove the "Pressure release” chamber
filter and check whether it is soiled/blocked
by packaging If necessary, clean the
chamber filter.
132
The steam sterilizer is overloaded or load
objects are arranged unfavourable.
Comply with the maximum permissible load
quantities (see
[
page 21]). The load may have no direct
contact to the pressure inlets or may not
cover them.
133
see event 124
136
a) The surrounding temperature of the steam
sterilizer is too hot.
b) The ventilation slots in the side walls are
clogged/blocked.
c) The steam sterilizer is installed. The
minimum clearance to the surrounding
surfaces has not been maintained.
d) The door was left open after loading or
unloading and hot steam has escaped from
the sterilization chamber.
Switch off the steam sterilizer and allow it to
cool for approx. 1 h.
a) Comply with the requirements at the
installation location and the maximum
ambient temperature (see Technical
Manual).
b) Clean the ventilation slots and remove any
objects covering them.
c) Maintain a minimum clearance to the
surrounding surfaces (see information in the
Technical Manual).
d) Always close the door after loading or
unloading.
175
The overheat control switch of the main
heating has tripped. This notification may be
issued in alternation with
E176: ACOUT 02
open
.
1. Switch off the steam sterilizer and press
the reset button overheat control on the fore
side of the steam sterilizer at the bottom
right-hand side (behind the cover) back in.
2. Acknowledge the malfunction message.
3. Switch off the steam sterilizer and back on
again and then perform an empty sterilization
run if necessary. Then the steam sterilizer is
ready for operation.
176
The overheat control switch of the main
heating has tripped. This notification may be
issued in alternation with
E175: ACOUT 01
open
.
1. Switch off the steam sterilizer and press
the reset button overheat control on the fore
side of the steam sterilizer at the bottom
right-hand side (behind the cover) back in.
2. Acknowledge the malfunction message.
3. Switch off the steam sterilizer and back on
again and then perform an empty sterilization
run if necessary. Then the steam sterilizer is
ready for operation.
182
The mains voltage is too low, poor building
voltage supply (e.g. undersized installation,
defective socket, multiple devices on a single
socket/fuse).
Arrange for an inspection of the building-side
installation (e.g. automatic circuit breaker)
and test the steam sterilizer at another
socket or on another circuit.
183
see event 124
185
see event 110
186
see event 110