11 Maintenance
66
The extent of stain accretion on the instruments is also dependant on the feed water used for steam gen-
eration.
Oiling the door spindle
Oil the door spindle
every two months
as follows:
1.
Clean the spindle with a non-fuzzing cloth.
2.
Place two drops of oil in the bushing on the door from the oil can included in the scope of delivery (art.
no. 27515).
Instructions for oiling the door spindle are printed on the inside of the door.
Maintenance
NOTICE
Continuing operation beyond the maintenance interval can result in malfunctions in the
device.
n
Maintenance should only be performed by trained and authorized service technicians or
stockist technicians.
n
Maintain the specified servicing intervals.
Regular maintenance is vital to ensure reliable operation and value retention of the steam sterilizer. All
function and safety-relevant components and electrical units must be checked during maintenance and re-
placed where necessary. Maintenance must be performed in accordance with the pertinent maintenance
instructions of the steam sterilizer.
Maintenance work is to be performed regularly in accordance with the 2000 program cycles but must be
performed after 24 months. The steam sterilizer will issue a maintenance message at the relevant time.