13 Malfunctions
76
Incident
Possible causes
What you can do
187
a) The chamber filter “Pressure release” is
blocked.
a) Remove the "Pressure release” chamber
filter (in the rear area of the sterilization
chamber below) and check whether it is e.g.
soiled/blocked by packaging. If necessary,
clean the chamber filter.
With connection to the building waste water
outflow:
The waste water cannot flow off.
b) The waste water hose is kinked or
installed sagging.
c) The siphon or the building-side waste
water line is blocked.
With connection to the building waste water
outflow:
b) Check the installation of the waste water
hose. This must be installed without kinking
or sagging and at a constant decline.
c) Check whether the building siphon is
blocked. Please note: If multiple devices are
operated simultaneously, we recommend the
installation of an additional siphon.
192
When using the internal storage tank:
a) The right-hand chamber of the internal
storage tank (feed water) must be sufficiently
filled with feed water for the pending rinsing
process.
a) Check the water level of the feed water in
the right-hand chamber of the internal
storage tank and refill with feed water if
necessary.
When using a MELAG water treatment unit:
b) The pending feed water procedure
requires sufficient supply with feed water.
b) Check whether the water inflow tap for the
water treatment unit is open. When using a
MELAdem 47, check whether the tap on the
pressure tank is open.
193
The left-hand chamber of the internal storage
tank (waste water) must be empty for the
pending rinsing process.
NOTICE! Danger of draining through the
emergency overflow.
Check the water level
and empty the left-hand chamber of the
internal storage tank (wastewater)
completely.
203
No log output options have been set.
Check the configuration in the menu
Settings
>
Logging
.
204
The internal log memory is full.
Output the log saved in the steam sterilizer
on any output medium or adapt the general
output options in the
Settings
>
Logging
menu.
207
see event 203
208
see event 204
211
see event 204
214
The steam sterilizer has not recognized the
CF card; it cannot be read, it is full or it is
damaged.
1. Check whether the CF card has been
inserted correctly (do not insert under
voltage).
2. Make sure that the CF card is not larger
than 4 GB.
3. Check whether the write-protection has
been set on the CF card by mistake.
4. Test the CF card on a computer.
5. Check whether the memory on the CF is
full. If the memory is full, transfer the log files
on the CF card to a computer and delete the
files on the CF card.
6. Transfer the log files on the CF card to a
computer and re-format the CF card in the
steam sterilizer.
7. The CF card is defective or incompatible.
It is possible that a non-MELAG CF card has
been used.
Please note: We recommend using only
original MELAG CF cards.