7 Sterilization
29
2.
Confirm with [ENTER]. Now you can display the
program run or information about your steam sterilizer
(e. g. serial number, device software version and
selected values).
Manual program abort
You can abort a current program in all phases. If you end a program before drying begins, the sterilization
material remains
unsterile
.
WARNING
Hot steam can be released from the device when opening the door after a program abort.
This could result in burns.
n
Use a tray lifter to remove the tray.
n
Never touch the sterilized equipment, the chamber or the door with bare hands. The
components are hot.
Program abort before the start of drying
WARNING
Danger of infection from early program abort
Aborting a program before the drying phase begins means that the load is unsterile. This
endangers the health of your patients and practice team.
n
If necessary, repack the load and repeat the sterilization for the sterilization material
affected.
Should you still wish to do so, proceed as follows to abort the program during drying:
1.
Press CANCEL on the action bar.