8 Logging
36
4.
Mark the editor and confirm with "OK".
Ê
You can then open files with this ending via a double-click in Windows Editor.
Using the label printer as an output medium
The use of a label printer facilitates batch traceability. Using the sterilization date, the storage duration,
batch number, user ID of the person approving the application for use, the steam sterilizer used and the
file name it is possible to assign the sterilized instruments to the patient and sterilization batch. Faultless
packages containing sterilized equipment are marked with labels after sterilization. As such, the precondi-
tions for correct "Approval" by the person conferred with the task of decontamination are given. All informa-
tion regarding the correct sterilization procedure can be attributed to the instruments used in patient re-
cords.
PLEASE NOTE
To facilitate easy assignation of a package marked with a label to a specific batch, the
sterilization log file name must not be changed.
Type, serial number
Sterilization date/Storage duration
Daily batches/no. total batches
User ID
File name
Storage period
MELAG 41 B+ 201741-B0112
05.04.2018 / 04.10.2018
Batch 03/00019
ID: 1001
Outputting text logs automatically after program end
(immediate output)
Should you wish to output the associated text and graphic logs (optional) on an output medium immediate-
ly after the end of a program, use option
Immediate output
. In its delivery state, the immediate output
of the text and graphic logs via the CF card after program end is activated.
If the output medium selected for this purpose has not been connected, the logs are saved in the internal
memory and a warning is issued. The steam sterilizer provides the option of outputting this log at the next
possible opportunity. Graphic logs cannot be saved in the internal log memory; they are lost. Further infor-
mation pertaining to the output of graphic logs is available in the section
Outputting graphic logs (optional)
[