7-70
Code
Cl.
Operational
message
Description
Input
64
R
DAC timeout
Automatic „HVOff“ after
the expiry of a defined
time frame without
commands from the PD-
measurement system
Restart measurement.
68
R
Unknown
version
Software version of the
TDS40/60 is not
compatible with the
software version of the
PD measurement system.
Contact service and
enquire about necessary
steps.
69
R
Task conflict
Software problem -
must not occur during operating.
70
R
Not connected
71
R
Watchdog failed
72
R
Card inserted
In the remote control
mode a SystemCard has
been inserted.
Remove the SystemCard!
73
R
Message lapse
Software problem -
must not occur during operating.
74
R
State mismatch
75
R
Unknown
request
76
R
Parameter error
77
R
Lost connection
Communication between
the TDS40/60 and the PD
measurement system has
been interrupted.
Potential interruption of
the cable connection or
crash of the PD software.
Solve the problem and
restart the system.