6-45
•
Printout
„Extended“:
- All
information listed under “Standard”
- Maximum test voltage, test voltage increase, step increment, rest period
of the ramp (ramp pause)
- Settings for logging and printing
- User profile, service / administrator mode (if applicable)
- Measured data (time, test voltage, current) of the charging period
•
Printout
„Detailed“:
- All information listed under “Extended”
- A bunch of measuring data logged right before the end of the test
For systems equipped with the logging option (see section 6.4), up to 4 header lines and 4
footer lines (e.g. the company name or a field for the signature of the operator) can be
specified.
These so called individual “
print templates
” can be imported into the system using a
SystemCard (see section 6.4.2). It depends on the settings of the SystemCard whether
these print templates are permanently stored in the test system configuration or are only
valid for the period of the cable test under way.
In order to test the layout of the individual header lines and footer lines you can initiate a
test printout even if no cable test has been performed before. For this purpose, you have
to change the print settings to “Text” or “System” first. Afterwards, you can initiate the test
printout from the start menu by pressing F1.
Depending on the print settings, the following information are included in the printout:
•
Printout
„Text“:
- Header (extended by the header from the SystemCard (if specified))
- System type and version
- Footer (extended by the footer from the SystemCard (if specified))
- Date / Time of the printout
•
Printout
„System“:
- All information listed under “
Text
”
- All information about the system configuration (options, software
versions)
- If necessary, all calibration data from the system
Содержание TDS40
Страница 2: ......
Страница 19: ...2 19 2 6 Testable Cable Capacitance in VLF Mode...
Страница 20: ...2 20 2 7 Cable Capacitance that Can be Diagnosed in DAC Mode TDS40 TDS60...
Страница 26: ......