8. A dialog box will appear with a warning that the voltage will lock and start at the target voltage level for all
subsequent tests, overriding the zero-volt interlock feature.
Fig 46:
Zero start override warning message.
9. Move the test fixture (or probes) to the second bar pairing (2–3).
10. Press the
Test (PTT)
button or footswitch. The analyzer will energize Lead 1 at the preset voltage and the
waveform will be displayed on the screen as the red wave. The reference is also displayed as the blue waveform.
A reference-to-bar EAR value will be displayed in the
Ref Bar EAR
field.
11. If the EAR percentage is greater than the tolerance, the field background will turn red indicating a failed test. If
the EAR percentage is under or equal to the tolerance, the background of the
Ref-Bar EAR
box will be white,
indicating a passing test.
12. When you are satisfied with the waveform, release the fixture’s test button or footswitch. The EAR percentage
will be saved and an inset bar graph will be displayed depicting the EAR percentage versus the reference bar’s
waveform. If the bar failed, the EAR bar on the graph will be red.
13. Up to three additional waveforms can be saved during the testing process. If a bar fails or is of interest, click
on one of the
Save
button in the
Save Waveforms
section at the top of the screen and that waveform will be
saved along with its bar number; otherwise, only the reference-to-bar EAR percentage for the bar will be saved.
14. Continue to move the test fixture to the next bar and repeat the process from step 10 for each bar until all bars
have been tested. A maximum of 1,024 bars can be tested.
15. If at any time before you save the full test results you want to retest a bar, you can enter that bar number
directly into the
Bar Control
field, or use the up and down arrows to navigate to the bar number, and retest as
needed. Retesting will overwrite the previous data for that bar.
16. The
Total Bars Tested
number gives you the number of the last bar tested. Add one to that number to resume
testing where you left off.
Using Power Packs with Baker AWA Testers
www.megger.com 49
Содержание PPX30
Страница 1: ...Baker Power Packs PPX30 PPX30A PPX40 User Guide...
Страница 2: ......
Страница 6: ...Notices iv Baker Power Packs User Guide 71 061RC EN V2 www megger com...
Страница 12: ...Baker Power Pack Introduction 4 Baker Power Packs User Guide 71 061RC EN V2 www megger com...
Страница 24: ...Baker Power Pack Setup 16 Baker Power Packs User Guide 71 061RC EN V2 www megger com...
Страница 28: ...Using Power Packs with Baker AWA Testers 20 Baker Power Packs User Guide 71 061RC EN V2 www megger com...
Страница 74: ...66 Baker Power Packs User Guide 71 061RC EN V2 www megger com...
Страница 78: ...70 Baker Power Packs User Guide 71 061RC EN V2 www megger com...
Страница 82: ...74 Baker Power Packs User Guide 71 061RC EN V2 www megger com...