Running the Manual Armature Surge Test Using the Baker AWA and PPX30A
NOTE: In most applications, you will need help from another person to complete at least the first part of this test
process: obtaining the reference waveform. One person will operate the fixture on the armature while the other
operates the tester.
CAUTION: Ensure that both operators clearly understand the process and how to use the software. The fixture/
probe operator must know how to properly use the ATF-5000 (or test probes) to avoid injury from electrical shock, or
damaging the test equipment or the unit under test.
1. For this example, a new Motor ID for the armature, along with a new Test ID, have been created. For your
application, ensure that you have selected the proper Motor ID and Test ID for your armature.
2. From the
Tests
tab, ensure that the selected Test ID has the Surge test turned
ON
and the section is labeled
“Surge Test Using PP185.”
3. Click on the test configuration button at the end of the
Surge Test
row to open the
Surge Test
setup window.
Because this is a manual test, it is run from the
Surge Test - Armature
window.
4. The armature operator will position the test fixture (or probes) on the first bar and the second bar (adjacent to
the right); bar 1 will be the reference bar. Bars are tested in a progressing sequence of 1–2, 2–3, 3–4, and so on.
5. When both operators are ready, the armature operator will hold down one of the test buttons on the ATF-5000
(or footswitch if using probes), then the AWA operator will ramp the voltage to the desired level.
6. After the first bar is tested and you are satisfied that you have a good reference waveform, the AWA operator
will click on the
Set Ref
button. This establishes the reference for the remainder of the test and sets the power
pack’s target voltage level.
7. In the following example, the target waveform is acquired from bar 1, so the
Set Ref
button is pressed.
Fig 45:
Acquiring the reference waveform for armature (low-impedance coil) testing.
Using Power Packs with Baker AWA Testers
48
Baker Power Packs User Guide 71-061RC EN V2
www.megger.com
Содержание PPX30
Страница 1: ...Baker Power Packs PPX30 PPX30A PPX40 User Guide...
Страница 2: ......
Страница 6: ...Notices iv Baker Power Packs User Guide 71 061RC EN V2 www megger com...
Страница 12: ...Baker Power Pack Introduction 4 Baker Power Packs User Guide 71 061RC EN V2 www megger com...
Страница 24: ...Baker Power Pack Setup 16 Baker Power Packs User Guide 71 061RC EN V2 www megger com...
Страница 28: ...Using Power Packs with Baker AWA Testers 20 Baker Power Packs User Guide 71 061RC EN V2 www megger com...
Страница 74: ...66 Baker Power Packs User Guide 71 061RC EN V2 www megger com...
Страница 78: ...70 Baker Power Packs User Guide 71 061RC EN V2 www megger com...
Страница 82: ...74 Baker Power Packs User Guide 71 061RC EN V2 www megger com...