Baker AWA to Power Pack Setup
1. Set the analyzer on top of the power pack and secure it with the safety strap.
2. Connect the short AC power cable between the PPX AC outlet and AWA AC inlet receptacles.
3. Connect the 25-pin interconnect cable to the two units. The cable is marked on each end. Be sure to plug in the
end marked HOST into the auxiliary port (AUX /PP) on the front of the Baker AWA-IV and the end marked 30kV
into the auxiliary port (AUX) on the front of the power pack.
4. Connect the long AC power cord from the AC receptacle on the side of the power pack to a properly grounded
AC source.
Fig 12:
Connect power pack to AC source and 25-pin interconnect cable.
Item
Description
1
25-pin interconnect cable—AWA and PPX front panel connectors
2
PPX AC outlet to AWA AC inlet receptacle
3
PPX AC inlet to properly grounded AC source
Baker Power Pack Setup
www.megger.com 19