Define Test Parameters for the Test ID
1. From the
Tests
tab, turn on the Resistance, Megohm, PI, HiPot, and Surge tests.
2. Open each test setup window to properly configure the tests. The
DC Tests
setup window is shown in the
example below.
Fig 15:
Configuring DC tests for a new Test ID.
Notice that all tests have been turned on, the desired test voltages entered, and the
Enable PP
box under the HiPot
column has been checked, indicating that the power pack will be used for the HiPot test.
You should also note that you can use the Dischg Time (min) fields to specify a discharge interval (in minutes) that will
stop all testing for the requested time period to allow for sufficient time for the motor to discharge. The feature is
more commonly used to discharge motors following DC HiPot testing.
CAUTION: If the windings are not allowed to completely discharge, a voltage can develop on the motor leads that
can be high enough to present a shock hazard to personnel.
Using Power Packs with Baker AWA Testers
www.megger.com 23
Содержание PPX30
Страница 1: ...Baker Power Packs PPX30 PPX30A PPX40 User Guide...
Страница 2: ......
Страница 6: ...Notices iv Baker Power Packs User Guide 71 061RC EN V2 www megger com...
Страница 12: ...Baker Power Pack Introduction 4 Baker Power Packs User Guide 71 061RC EN V2 www megger com...
Страница 24: ...Baker Power Pack Setup 16 Baker Power Packs User Guide 71 061RC EN V2 www megger com...
Страница 28: ...Using Power Packs with Baker AWA Testers 20 Baker Power Packs User Guide 71 061RC EN V2 www megger com...
Страница 74: ...66 Baker Power Packs User Guide 71 061RC EN V2 www megger com...
Страница 78: ...70 Baker Power Packs User Guide 71 061RC EN V2 www megger com...
Страница 82: ...74 Baker Power Packs User Guide 71 061RC EN V2 www megger com...