37
When Concurrent Mode of testing is selected, the user is required to select how many taps is desired to
test simultaneously
However if the Non Concurrent Mode of testing is selected, the user is required to select which
individual tap is desired
Hitting the select tap button will activate the tap selection dialog
After either selecting the number of taps or the individual tap selection, the Winding Resistance test is
begun by selecting the blue play button.
4.9 Insulation Test
Selection of
Insulation Test button from the test configuration screen will bring the user to
the following Insulation Test screen