
41
Step 1
Raise the stretcher as
high as you can. However,
do not exceed the height of the slider.
Entering the Ambulance in the Emergency Mode
WARNING
Before entering the ambulance, fully extend the arm of the loading system and
make sure that it automatically locks against return motion. Follow the instructions
in the VIVERA FASTENER user manual.
FRONT OF THE MONOBLOCK ON THE SLIDER
SLIDER
LOADING
SYSTEM ARM
OPTIMAL RELATIVE HEIGHT POSITION
OF THE STRETCHER TO THE SLIDER
Step 2
Move or lift the front of the monoblock
on the slider. Try to keep the monoblock in
the longitudinal plane of the loading system.
If the bed is tilted too much, the back
of the monoblock must be raised and placed
on the arm of the loading system. Push
the monoblock further inwards the
ambulance until it is automatically secured
against backward movement.
Make sure this has happened by pulling
the monoblock slightly out
of the ambulance.
Содержание M301V01
Страница 1: ...Rev A 25 6 2021 VIVERA Monoblock M301 M301V01 Operation and Maintenance Manual...
Страница 12: ...12 Modification M01 Modification M04 0 50 30 30 0 30 0 90 0 30 0 90...
Страница 13: ...13 Modification M05 Modification M08 0 50 30 30 0 30 0 90 0 30 0 90...
Страница 54: ...54 Maintenance Records Date Description of maintenance performed Performed by...
Страница 55: ...2021 MEDIROL s r o All rights reserved The MEDIROL logo is a registered trademark of MEDIROL s r o...