23
Height Positioning
Use height positioning only according to the intended use of this product.
The product allows a very wide range of height positions. When transporting outside
the ambulance, set a suitable height for safe handling and good visual control
of the patient and the surrounding terrain. Use the lowest position only for loading
or unloading the patient. When loading into the ambulance with a higher floor, raise
the chassis to the appropriate height just before loading.
The “+” button activates the upward movement while the “-” button activates
the downward movement of the bed.
When moving upwards, it automatically stops at the default loading height. By pressing
the “+” button again, the stretcher can also be raised above this position.
NOTE
Shortly after pressing the “-” button, the stretcher is slightly raised upwards. In this
way, the system optimizes the control according to the load. This is not a malfunction.
WARNING
Never activate height positioning at an angle of more than 30° from the horizontal
level. There is a risk of damage to the product.
WARNING
Observe the maximum prescribed duty cycle specified in the chapter Technical
Specifications (p. 10). There is a risk of the system overheating.
Содержание M301V01
Страница 1: ...Rev A 25 6 2021 VIVERA Monoblock M301 M301V01 Operation and Maintenance Manual...
Страница 12: ...12 Modification M01 Modification M04 0 50 30 30 0 30 0 90 0 30 0 90...
Страница 13: ...13 Modification M05 Modification M08 0 50 30 30 0 30 0 90 0 30 0 90...
Страница 54: ...54 Maintenance Records Date Description of maintenance performed Performed by...
Страница 55: ...2021 MEDIROL s r o All rights reserved The MEDIROL logo is a registered trademark of MEDIROL s r o...