19
2.
1.
Exiting the Ambulance
System Activation
The system is activated by briefly pressing the “-” or “+” button. Running is indicated
by lights on the indicator panel.
The system automatically shuts off after 15 minutes of inactivity.
FIXING LEVER
LIGHT ON THE INDICATOR PANEL AFTER PRESSING THE BUTTON
Monoblock Operation
WARNING
Securing against backward movement after the monoblock has been pulled out
from the ambulance partially prevents an unwanted backward movement towards
the ambulance. However, this is not sufficient protection against movement due to
external forces in real conditions. Never leave the patient unattended.
Step 1
Press the fixing lever and move the
monoblock completely out of the ambulance.
After moving the monoblock to the end
position, the auxiliary spring is pushed,
which prevents the unintended return
movement towards the ambulance. When
you push the monoblock slightly towards the
ambulance, you will feel a slight resistance
of the spring, which will make you sure that
it is in the limit position. Before the next
step, always make sure that the monoblock
is completely in the limit position.