The medi pcs brio system

1) Pressure Control Unit (PCU)

The PCU 

(2.1.1)

 is a programmable pneumatic compressor with 

two connector outlets. Each connector has ten outflow ports 

into which the garment hoses plug. Air passes through the 

hoses, delivering treatment through the sequential inflation and 

deflation of up to eight air chambers in the garments, or sixteen 

chambers total if two garments are being used. By programming 

a treatment program using the touchscreen Graphical User 

Interface (GUI), targeted pressure is delivered to the chambers 

and assists in moving excess fluid out of affected limb(s). 

2) Blocking Plate

The blocking plate 

(2.1.2)

 is used to cover an open connector 

outlet. If the patient is using only one garment, you must install 

the blocking plate in the open connector outlet. The PCU will not 

operate properly if there is an uncovered, open connector outlet, 

resulting in an error. If the patient is using two garments at the 

same time during treatment, the blocking plate can be set aside 

but should not be discarded. 

3) garment Connector

The garment connector 

(2.1.3)

 attaches the garment to the PCU. 

The number of open/closed ports will vary based on compression 

garment model. See Chapter 4 for compression garment sizing 

and application information.

4) Power Supply

The power supply 

(2.1.4)

 connects the PCU to an electrical outlet.

Step 1: When your medi pcs brio arrives, carefully unpack the 

contents and ensure that all equipment required to begin 

operation is present. 

Step 2: Place the PCU on a flat, stable surface where the patient 

will easily be able to operate the controls during treatment. 

(2.2.1) 

The surface should be free of miscellaneous items such 

as loose paper, fabric, or tablecloths that could cause the air 

outflow ports to become blocked during treatment.

Step 3: Attach the two parts of the power supply to each other. 

(2.2.2)

 Locate the power supply jack on the back of the PCU and 

plug the circular end of the power supply into the PCU 

(2.2.3)

.

Step 4: Plug the opposite end of the power supply into an outlet.

The medi pcs brio system

Components

 System assembly

2.1

2.2

pressure control unit (PCU)

garment connector

blocking plate

power supply

Components:

• Operator’s Manual

• Garment Application & 

   Maintenance Guide

• Patient Quick Start Guide

• Pressure Control Unit (PCU)

• Blocking Plate 

• Power Supply

• Garment(s) 

(the garment(s) received will 

vary depending 

on individual patient orders)

NOTE: Garments should not be connected to the PCU until after 

garments have been applied to the appropriate limbs to reduce 

the risk of accidentally pulling the PCU off its resting surface.

(2.1.1)

(2.1.2)

(2.1.4)

(2.1.3)

(2.2.1)

(2.2.3)

(2.2.2)

Содержание Pcs brio

Страница 1: ...medi pcsbrio medi pcs brio pneumatic compression system Operator s Manual...

Страница 2: ...2...

Страница 3: ...session 6 4 Turning off the PCU 6 5 Error screen 7 1 Storing the medi pcs brio 7 2 Cleaning the medi pcs brio 8 1 Troubleshooting 8 2 Warranty return policy 8 3 Safety warnings precautions 8 4 Labels...

Страница 4: ...stingpulmonary edema existingpulmonaryembolism inflammatoryphlebitis severecongestivecardiacfailure severeatherosclerosisor otherischemicvasculardiseases anycircumstancewhere increasedvenousandlymphat...

Страница 5: ...ression garment sizing and application information 4 Power Supply The power supply 2 1 4 connects the PCU to an electrical outlet Step 1 When your medi pcs brio arrives carefully unpack the contents a...

Страница 6: ...he garment going up the garment from distal to proximal The step value setting can be changed in 1 mmHg increments ranging from 1 60 mmHg However the step value range is limited such that no chamber w...

Страница 7: ...distal pressure is set to 50 mmHg with a step value of 3 mmHg Chamber 1 2 3 4 5 6 mmHg 50 47 44 41 38 35 Example 2 An 8 chamber garment whole leg or arm where the distal pressure is set to 80 mmHg wit...

Страница 8: ...e medi pcs brio is now ready to operate Please review the settings displayed in the Treatment Summary screen to ensure that all settings have been entered correctly If you must unlock or change the tr...

Страница 9: ..._____________________________ Language ______________________________ Treatment time ____________________________ Distal pressure ______________________________ Access level __________________________...

Страница 10: ...cm Lower Leg Sizing Measurements T i p o f M i d d l e F i n g e r t o A x i l l a C G Floor to Knee Crease B C Floor to Gluteal Fold C G B Arm Garment Sizing Measurements Size Item Leg Length Circumf...

Страница 11: ...before applying If desired apply a liner over the limb before applying the garment This will help keep the garment clean during use Loose unrestrictive clothing such as sweatpants or cotton t shirts m...

Страница 12: ...arment either slide your limb out of the garment or undo the hook and loop material and remove your limb It is recommended to keep the hook and loop material fastened even when the compression garment...

Страница 13: ...e latch mechanism and pulling the connector out The blocking port can remain in place for future treatment sessions Step 1 Locate the garment connectors at the end of the hoses attached to the garment...

Страница 14: ...appear 6 4 The PCU will begin the fill to fit process after which the programmed treatment cycle will begin The Treatment Running screen will show treatment time remaining current active chamber in us...

Страница 15: ...t Summary Screen Press the power switch on the back of the PCU to turn off the medi pcs brio Please note that the PCU inflation deflation valves close while the PCU is turned off If a garment is conne...

Страница 16: ...sure that the power supply is unplugged before cleaning the PCU Gently wipe the PCU with a slightly damp cloth Do not allow any moisture or liquid to enter the device If additional cleaning is require...

Страница 17: ...d any rapid movements or drastic posture changes during the first 5 10 minutes of treatment 2 Verify that the correct treatment settings have been selected by referring to the Treatment Summary screen...

Страница 18: ...e or service the device while in use To completely eliminate power disconnect the power supply from the Pressure Control Unit or from the wall outlet Warning Electric Shock Hazard Unplug power supply...

Страница 19: ...nce pain or notice your condition worsening contact your physician Do not change your treatment settings without consulting your physician Warning Parental Supervision Children under the age of 18 sho...

Страница 20: ...ing the safety requirements specified for Class II equipment according to IEC 61140 Non ionizing electro magnetic radiation ISO 7000 IEC 60417 Graphical symbols for use on equipment 5140 To indicate o...

Страница 21: ...47 Part15 Subpart B ICES 001 ISSUE 4 AS NZS CISPR 11 JIS T 0601 1 2 ABNT NBR IEC 60601 1 2 LEGAL MANUFACTURER medi USA LP 6481 Franz Warner Parkway Whitsett NC 27377 www mediusa com Made in USA Tempe...

Страница 22: ...perating normally WARNING Use of accessories transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or...

Страница 23: ...cle At 0 45 90 135 180 225 270 and 315 0 UT 1 cycle and 70 UT 25 30 cycles h Single phase at 0 0 UT 250 300 cycle 0 UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 0 UT 1 cycle and 70 UT 25 30 cycles...

Страница 24: ...ncy range b Interference may occur in the vicinity of equipment marked with the following symbol Guidance and Manufacturer s Declaration Electromagnetic Immunity NOTE 1 At 80 MHz and 800 MHz the highe...

Страница 25: ...o any part of the medi pcs including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance E 6 d P...

Страница 26: ...nd mobile RF communications equipment transmitters and the medi pcs as recommended below according to the maximum output power of the communications equipment For transmitters rated at a maximum outpu...

Страница 27: ...27 For other language translations please visit www mediusa com...

Страница 28: ...medi USA 6481 Franz Warner Parkway Whitsett NC 27377 info mediusa com T 800 633 6334 F 888 570 4554 mediusa com PBO157 Rev A...

Отзывы: