background image

24

IMMUNITY TEST

IEC 60601 Test Level

Compliance Level

Electromagnetic 

Environment - Guidance

ECONDUCTED RF

IEC 61000-4-6

RADIATED RF

IEC 61000-4-3

3 Vrms

0,15 MHz – 80 MHz

6 Vrms in ISM and amateur 

radio bands between 0,15 

MHz and 80 MHz 

80 % AM at 1 kHz

10 V/m

80 MHz to 2.7 GHz

3 Vrms

0,15 MHz – 80 MHz

6 Vrms in ISM and amateur 

radio bands between 0,15 

MHz and 80 MHz 

80 % AM at 1 kHz

10 V/m

80 MHz to 2.7 GHz

Portable and mobile RF 

communications 

equipment should be used no 

closer to any part 

of the medi pcs, including 

cables, than the 

recommended separation 

distance calculated 

from the equation applicable 

to the frequency of 

the transmitter.

Recommended separation 

distance

d = [3.5/10] √P 80 MHz to 800 

MHz

d = [7/10] √P 800 MHz to 2.7 

GHz

where P is the maximum 

output power rating of the 

transmitter in watts (W) 

according to the transmitter 

manufacturer and d is the 

recommended separation 

distance in meters (m). Field 

strengths from fixed RF

transmitters, as determined 

by an electromagnetic site 

survey, a should be less 

than the compliance level 

in each frequency range. b 

Interference may occur in the 

vicinity of equipment marked 

with the following symbol:

Guidance and Manufacturer’s Declaration – Electromagnetic Immunity

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 

reflection from structures, objects and people.

The medi pcs is intended for use in the electromagnetic environment specified below.  The 

customer or the user of the medi pcs should assure that it is used in such an environment.

• Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile 

radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.  

To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be 

considered. If the measured field strength in the location in which the medi pcs is used exceeds the applicable RF 

compliance level above, the medi pcs should be observed to verify normal operation. If abnormal performance is observed, 

additional measures may be necessary, such as re-orienting or relocating the medi pcs.

Electromagnetic Capability

Содержание Pcs brio

Страница 1: ...medi pcsbrio medi pcs brio pneumatic compression system Operator s Manual...

Страница 2: ...2...

Страница 3: ...session 6 4 Turning off the PCU 6 5 Error screen 7 1 Storing the medi pcs brio 7 2 Cleaning the medi pcs brio 8 1 Troubleshooting 8 2 Warranty return policy 8 3 Safety warnings precautions 8 4 Labels...

Страница 4: ...stingpulmonary edema existingpulmonaryembolism inflammatoryphlebitis severecongestivecardiacfailure severeatherosclerosisor otherischemicvasculardiseases anycircumstancewhere increasedvenousandlymphat...

Страница 5: ...ression garment sizing and application information 4 Power Supply The power supply 2 1 4 connects the PCU to an electrical outlet Step 1 When your medi pcs brio arrives carefully unpack the contents a...

Страница 6: ...he garment going up the garment from distal to proximal The step value setting can be changed in 1 mmHg increments ranging from 1 60 mmHg However the step value range is limited such that no chamber w...

Страница 7: ...distal pressure is set to 50 mmHg with a step value of 3 mmHg Chamber 1 2 3 4 5 6 mmHg 50 47 44 41 38 35 Example 2 An 8 chamber garment whole leg or arm where the distal pressure is set to 80 mmHg wit...

Страница 8: ...e medi pcs brio is now ready to operate Please review the settings displayed in the Treatment Summary screen to ensure that all settings have been entered correctly If you must unlock or change the tr...

Страница 9: ..._____________________________ Language ______________________________ Treatment time ____________________________ Distal pressure ______________________________ Access level __________________________...

Страница 10: ...cm Lower Leg Sizing Measurements T i p o f M i d d l e F i n g e r t o A x i l l a C G Floor to Knee Crease B C Floor to Gluteal Fold C G B Arm Garment Sizing Measurements Size Item Leg Length Circumf...

Страница 11: ...before applying If desired apply a liner over the limb before applying the garment This will help keep the garment clean during use Loose unrestrictive clothing such as sweatpants or cotton t shirts m...

Страница 12: ...arment either slide your limb out of the garment or undo the hook and loop material and remove your limb It is recommended to keep the hook and loop material fastened even when the compression garment...

Страница 13: ...e latch mechanism and pulling the connector out The blocking port can remain in place for future treatment sessions Step 1 Locate the garment connectors at the end of the hoses attached to the garment...

Страница 14: ...appear 6 4 The PCU will begin the fill to fit process after which the programmed treatment cycle will begin The Treatment Running screen will show treatment time remaining current active chamber in us...

Страница 15: ...t Summary Screen Press the power switch on the back of the PCU to turn off the medi pcs brio Please note that the PCU inflation deflation valves close while the PCU is turned off If a garment is conne...

Страница 16: ...sure that the power supply is unplugged before cleaning the PCU Gently wipe the PCU with a slightly damp cloth Do not allow any moisture or liquid to enter the device If additional cleaning is require...

Страница 17: ...d any rapid movements or drastic posture changes during the first 5 10 minutes of treatment 2 Verify that the correct treatment settings have been selected by referring to the Treatment Summary screen...

Страница 18: ...e or service the device while in use To completely eliminate power disconnect the power supply from the Pressure Control Unit or from the wall outlet Warning Electric Shock Hazard Unplug power supply...

Страница 19: ...nce pain or notice your condition worsening contact your physician Do not change your treatment settings without consulting your physician Warning Parental Supervision Children under the age of 18 sho...

Страница 20: ...ing the safety requirements specified for Class II equipment according to IEC 61140 Non ionizing electro magnetic radiation ISO 7000 IEC 60417 Graphical symbols for use on equipment 5140 To indicate o...

Страница 21: ...47 Part15 Subpart B ICES 001 ISSUE 4 AS NZS CISPR 11 JIS T 0601 1 2 ABNT NBR IEC 60601 1 2 LEGAL MANUFACTURER medi USA LP 6481 Franz Warner Parkway Whitsett NC 27377 www mediusa com Made in USA Tempe...

Страница 22: ...perating normally WARNING Use of accessories transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or...

Страница 23: ...cle At 0 45 90 135 180 225 270 and 315 0 UT 1 cycle and 70 UT 25 30 cycles h Single phase at 0 0 UT 250 300 cycle 0 UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 0 UT 1 cycle and 70 UT 25 30 cycles...

Страница 24: ...ncy range b Interference may occur in the vicinity of equipment marked with the following symbol Guidance and Manufacturer s Declaration Electromagnetic Immunity NOTE 1 At 80 MHz and 800 MHz the highe...

Страница 25: ...o any part of the medi pcs including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance E 6 d P...

Страница 26: ...nd mobile RF communications equipment transmitters and the medi pcs as recommended below according to the maximum output power of the communications equipment For transmitters rated at a maximum outpu...

Страница 27: ...27 For other language translations please visit www mediusa com...

Страница 28: ...medi USA 6481 Franz Warner Parkway Whitsett NC 27377 info mediusa com T 800 633 6334 F 888 570 4554 mediusa com PBO157 Rev A...

Отзывы: