Step 1: Treatment Summary Screen 

(3.2)

This screen shows all the treatment parameters for the 

program currently stored in memory. There are two buttons 

on the Treatment Summary screen:

  

 

 

: This will take you to the main settings  

 

 

screen where you can adjust all the PCU settings. 

 

If   the Settings button is not visible, proceed to  

 

 Section 

3.3 to unlock the PCU. 

  

 

: This will run the current treatment settings.

Press 

  to begin programming treatment settings.

Step 2: Select Language 

(3.3)

. Select one of the four 

language options, English, French, Spanish, or Portuguese 

by pressing  

  until the desired language is selected.

Step 3: Set Distal Pressure 

(3.4)

The Distal Pressure is 

the amount of compression that will be applied by the 

most distal chamber of the garment (at the hand or foot, 

depending on the limb being treated). The Distal Pressure 

can be set from 20-80 mmHg. Press  

  to increase                 

the distal pressure in increments of 1 mmHg. Press

 

 

to 

decrease the distal pressure in increments of 1 mmHg.

Step 4: Set Step Value 

(3.5)

. The step value is the increment 

by which the compression level will decrease between 

each chamber of the garment going up the garment from 

distal to proximal. The step value setting can be changed in 

1 mmHg increments ranging from 1–60 mmHg. However, 

the step value range is limited such that no chamber will 

ever allow a pressure of less than 20 mmHg. All subsequent 

chambers will also be filled to the 20 mmHg minimum 

pressure level. (See Example 2.)

Powering on

3.1

Initial setup

3.2

3

Programming treatment settings

Treatment settings should only be programmed by Qualified Medical Professionals. Patients should refer to the separate

 

Patient Quick Start Guide for instructions on running a treatment session.
Before using the medi pcs brio for the first time, ensure that a treatment area has been established for the patient so that he or 

she can easily access the PCU  Graphical User Interface (GUI) at all times.

Turn on the PCU by pressing the power 

switch on the back side of the PCU. 

Once on, the Startup Venting screen 

will be displayed while the chambers 

vent

 (3.1)

.

NOTE: When programming treatment settings, press 

  to save your settings and return to the previous 

treatment setting. Press 

  to save your settings and 

move onto the next treatment setting. When adjusting 

any of the settings, hold 

   or  

 for several seconds 

continuously to adjust values more quickly

.

(3.1)

(3.2)

(3.4)

(3.3)

Programming treatment settings

Содержание Pcs brio

Страница 1: ...medi pcsbrio medi pcs brio pneumatic compression system Operator s Manual...

Страница 2: ...2...

Страница 3: ...session 6 4 Turning off the PCU 6 5 Error screen 7 1 Storing the medi pcs brio 7 2 Cleaning the medi pcs brio 8 1 Troubleshooting 8 2 Warranty return policy 8 3 Safety warnings precautions 8 4 Labels...

Страница 4: ...stingpulmonary edema existingpulmonaryembolism inflammatoryphlebitis severecongestivecardiacfailure severeatherosclerosisor otherischemicvasculardiseases anycircumstancewhere increasedvenousandlymphat...

Страница 5: ...ression garment sizing and application information 4 Power Supply The power supply 2 1 4 connects the PCU to an electrical outlet Step 1 When your medi pcs brio arrives carefully unpack the contents a...

Страница 6: ...he garment going up the garment from distal to proximal The step value setting can be changed in 1 mmHg increments ranging from 1 60 mmHg However the step value range is limited such that no chamber w...

Страница 7: ...distal pressure is set to 50 mmHg with a step value of 3 mmHg Chamber 1 2 3 4 5 6 mmHg 50 47 44 41 38 35 Example 2 An 8 chamber garment whole leg or arm where the distal pressure is set to 80 mmHg wit...

Страница 8: ...e medi pcs brio is now ready to operate Please review the settings displayed in the Treatment Summary screen to ensure that all settings have been entered correctly If you must unlock or change the tr...

Страница 9: ..._____________________________ Language ______________________________ Treatment time ____________________________ Distal pressure ______________________________ Access level __________________________...

Страница 10: ...cm Lower Leg Sizing Measurements T i p o f M i d d l e F i n g e r t o A x i l l a C G Floor to Knee Crease B C Floor to Gluteal Fold C G B Arm Garment Sizing Measurements Size Item Leg Length Circumf...

Страница 11: ...before applying If desired apply a liner over the limb before applying the garment This will help keep the garment clean during use Loose unrestrictive clothing such as sweatpants or cotton t shirts m...

Страница 12: ...arment either slide your limb out of the garment or undo the hook and loop material and remove your limb It is recommended to keep the hook and loop material fastened even when the compression garment...

Страница 13: ...e latch mechanism and pulling the connector out The blocking port can remain in place for future treatment sessions Step 1 Locate the garment connectors at the end of the hoses attached to the garment...

Страница 14: ...appear 6 4 The PCU will begin the fill to fit process after which the programmed treatment cycle will begin The Treatment Running screen will show treatment time remaining current active chamber in us...

Страница 15: ...t Summary Screen Press the power switch on the back of the PCU to turn off the medi pcs brio Please note that the PCU inflation deflation valves close while the PCU is turned off If a garment is conne...

Страница 16: ...sure that the power supply is unplugged before cleaning the PCU Gently wipe the PCU with a slightly damp cloth Do not allow any moisture or liquid to enter the device If additional cleaning is require...

Страница 17: ...d any rapid movements or drastic posture changes during the first 5 10 minutes of treatment 2 Verify that the correct treatment settings have been selected by referring to the Treatment Summary screen...

Страница 18: ...e or service the device while in use To completely eliminate power disconnect the power supply from the Pressure Control Unit or from the wall outlet Warning Electric Shock Hazard Unplug power supply...

Страница 19: ...nce pain or notice your condition worsening contact your physician Do not change your treatment settings without consulting your physician Warning Parental Supervision Children under the age of 18 sho...

Страница 20: ...ing the safety requirements specified for Class II equipment according to IEC 61140 Non ionizing electro magnetic radiation ISO 7000 IEC 60417 Graphical symbols for use on equipment 5140 To indicate o...

Страница 21: ...47 Part15 Subpart B ICES 001 ISSUE 4 AS NZS CISPR 11 JIS T 0601 1 2 ABNT NBR IEC 60601 1 2 LEGAL MANUFACTURER medi USA LP 6481 Franz Warner Parkway Whitsett NC 27377 www mediusa com Made in USA Tempe...

Страница 22: ...perating normally WARNING Use of accessories transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or...

Страница 23: ...cle At 0 45 90 135 180 225 270 and 315 0 UT 1 cycle and 70 UT 25 30 cycles h Single phase at 0 0 UT 250 300 cycle 0 UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 0 UT 1 cycle and 70 UT 25 30 cycles...

Страница 24: ...ncy range b Interference may occur in the vicinity of equipment marked with the following symbol Guidance and Manufacturer s Declaration Electromagnetic Immunity NOTE 1 At 80 MHz and 800 MHz the highe...

Страница 25: ...o any part of the medi pcs including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance E 6 d P...

Страница 26: ...nd mobile RF communications equipment transmitters and the medi pcs as recommended below according to the maximum output power of the communications equipment For transmitters rated at a maximum outpu...

Страница 27: ...27 For other language translations please visit www mediusa com...

Страница 28: ...medi USA 6481 Franz Warner Parkway Whitsett NC 27377 info mediusa com T 800 633 6334 F 888 570 4554 mediusa com PBO157 Rev A...

Отзывы: