16
Caring for your medi pcs brio system
Storing your system
Cleaning your system
7.1
7.2
7
Caring for your medi pcs brio
After treatment is complete and the PCU is turned off, unplug the power supply cord from the power outlet and from the
PCU. Disconnect the garments from the PCU.
Store the garments by first loosely coiling the hoses and garment connectors. Avoid kinking or crushing the hoses. Lay the
garments flat in a cool, dry place for storage.
Store the PCU in a cool, dry place, avoiding direct sunlight. Keep it away from excessive heat or cold, and keep away from
children and pets.
PCU: Ensure that the power supply is unplugged before cleaning the PCU. Gently wipe the PCU with a slightly damp cloth.
Do not allow any moisture or liquid to enter the device. If additional cleaning is required, the PCU can be wiped with a
light application of hospital-grade disinfectant, such as 70% Isopropyl Alcohol solution.
Garments: Ensure that the garments are disconnected from the PCU before cleaning them. Gently wipe the garment
using a slightly damp cloth, moistened with a mild anti-bacterial soap or 70% Isopropyl Alcohol solution. Do not allow
any moisture or liquid to enter the air inlets of the garment connectors. After wiping with a damp cloth, thoroughly dry
the garment using a soft towel or cloth and allow to air dry fully. Do not bleach, iron, or machine wash or machine dry the
garments.