13
Connecting the garments to the pressure control unit (PCU)
Step 1: Locate the garment connector at the end
of the hoses attached to the garment
(5.1)
. The
number of open ports will vary based on the
garment model.
Step 2: Insert the garment connector into the open
port on the front of the PCU
(5.2)
. The latching
mechanism on top of the garment connector should
face up and will firmly lock into position when
inserted properly into the PCU port. The garment
connector connects with minimal amount of force
and should never be jammed or forced into the PCU
port. The garment connector will not connect if
inserted upside down or sideways.
Step 3: Insert the blocking plate into the remaining
open port on the front of the PCU and ensure that it
is secure
(5.3)
. If needed, the garment connector and
blocking plate can be switched between ports. The
compression output is the same for both ports and
is not left- or right-dependent.
After treatment, disconnect the garment
connector from the PCU by pressing down on the
latch mechanism and pulling the connector out.
The blocking port can remain in place for future
treatment sessions.
Step 1: Locate the garment connectors at the end of
the hoses attached to the garments
(5.4)
.
Step 2: Insert the garment connector into the
open port on the front of the PCU
(5.5)
. The latching
mechanism on top of the garment connector should
face up and will firmly lock into position when
inserted properly into the PCU port. The garment
connector connects with minimal amount of force
and should never be jammed or forced into the PCU
port. The garment connector will not be able to
connect if inserted upside down or sideways.
Step 3: Insert the second garment connector into
the second port of the PCU using the same method
applied for the first garment connector
(5.6)
.
The blocking plate is not needed and can be set aside
for later use if running unilateral treatment.
After treatment, disconnect the garment connectors
from the PCU by pressing down on the latch
mechanism and pulling the connectors out.
Connecting the garments to the pressure control unit (PCU)
Unilateral treatment (treatment with one garment)
Bilateral treatment (treatment with two garments)
5.1
5.2
(5.1)
(5.4)
(5.2)
(5.5)
(5.3)
(5.6)
5