Appendix
89
Appendix
Ⅱ
Checklist for Technical Safety Check (TSC)-Every 24
months
Hospital:
Department:
Contact person:
Phone:
Model:
SN:
Visual Inspection
Electrical Safety
Function Inspection
□
Cleanliness, damage
completeness, and faults
affecting safety, damage and
readability of the label.
Particularly:
□
Electrical Safety Test
Electrical safety test is performed to detect
hidden electrical dangers. If these hidden
dangers are not detected in time, they may
cause injury of patients or operators.
A safety analyzer is used as the electrical
safety test device. Service personnel must
know how to use the safety analyzer.
The electrical safety test is performed
according to IEC/EN60601-1.
The electrical safety test must be
performed
in
normal
environment
(including temperature, humidity, and
atmospheric pressure).
The electrical safety test provided in this
section takes the 601 safety analyzer as an
example. The test performed using the
safety analyzers of other brands may differ
from the test in this section.
If the product fails to pass the electrical
safety test, contact the after-sales service
personnel of Medcaptain.
□
Accuracy Check method:
Select a infusion set, set the
infusion rate to 60 ml/h, fill
the infusion set with distilled
water, and lead the extension
tube on the set to the
measuring cylinder. Start the
infusion, stop the infusion 10
minutes later, and check the
liquid
volume
in
the
measuring
cylinder.
The
infusion rate passes the test if
the liquid volume falls within
the range of 9.5-10.5 ml