Sensor Information
27
l)
Fluid/patient side pressure test: Press the pressure sensor to check whether the pressure value changes.
If the pressure value does not change, the pressure sensor may be faulty.
m)
Pump door Hall signal: Check whether
1
is displayed when the door is open and
0
is displayed when
the door is closed. If the displayed value is incorrect, check whether the electromagnet is properly
installed and whether the Hall switch functions properly.
n)
Number of drops: Check whether the actual number of drops is consistent with that detected by the
drop sensor with naked eyes. If there is a large difference between them, check whether the drop
sensor is properly installed and whether the drop sensor is faulty.