Disassembly and Assembly
Confidential Information of MEDCAPTAIN MEDICAL TECHNOLOGY CO., LTD. 57 / 93
3
HP-60 motor speed measuring PCBA
1
4
Cross recessed pan head, spring washer,
and flat washer combination screw
M2x6 GB 9074.4
1
1.
Remove the connection cable (7008000021) of the pump finger position detection board.
2.
Screw out the cross recessed pan head, spring washer, and flat washer combination screw M2x6
GB 9074.4 (1), and take down the HP-60 pump finger position detection PCBA (2).
3.
Remove the connection cable (7000000025) of the HP-60 motor speed measuring board.
4.
Screw out the cross recessed pan head, spring washer, and flat washer combination screw M2x6
GB 9074.4 (4), and take down the HP-60 motor speed measuring PCBA (3).
1.5
Dismantling the Sliding Block and HP-60 Door Potentiometer Assembly
No.
Part Name
Quantity