Disassembly and Assembly
Confidential Information of MEDCAPTAIN MEDICAL TECHNOLOGY CO., LTD. 69 / 93
No.
Part Name
Quantity
1
HP-60 veneer
1
1.
Use tweezers to strip off the veneer (1) from the seam between the veneer and shell.
1.15
Dismantling the HP-60 Pump Door Board’s Connection FFC
No.
Part Name
Quantity
1
HP-60 pump door hole plug
1
2
PT2.3*6 cross recessed pan head drilling
PT thread tapping screw
1
3
HP-60 pump door board connection FFC
1
4
HP-60 bottom shell FPC cover
1
1.
Use tweezers to remove the HP-60 pump door hole plug (1).
2.
Remove the PT2.3*6 cross recessed pan head drilling PT thread tapping screws (2).
3.
Take down the HP-60 bottom shell FPC cover (4).
4.
Disconnect one end of the HP-60 pump door board’s connection FFC (3).
No.
Part Name
Quantity
Содержание HP-60
Страница 30: ...Maintenance 30 Please follow local laws to dispose the old battery...
Страница 35: ...Calibration 35 Schematic diagram...