Disassembly and Assembly
Confidential Information of MEDCAPTAIN MEDICAL TECHNOLOGY CO., LTD. 74 / 93
3
HP-60 fastener seat
1
4
HP-60 fastener
1
5
Shifting block plate pressure spring
1
1.
Take down the shifting block plate (1). The shifting block plate is connected through the fastener.
Be careful not to break off the fastener when taking down the shifting block plate.
2.
Remove the two PT2.3*6 cross recessed pan head drilling PT thread tapping screws.
3.
Take down the fastener seat (3) together with the fastener (4) and shifting block plate pressure
spring (5).
1.19
Assembly
Perform assembly in the reversed order of the disassembly procedure. When performing assembly, note
the following points:
1.
The screw types and specifications must be correct.
2.
All parts and components must be properly installed.
3.
For the parts where lubricating grease has been applied but already erased, the operator must
reapply lubricating grease on these parts.
4.
When installing the main PCBA, do not forget to install the main board shielding case and main
board protection plate.
5.
When installing the main motor, pay attention to the direction of the outgoing line.
6.
After installing the main motor and reduction gear assembly, manually make the gear rotate for
over one circle to ensure that the gear assembly rotates steadily and smoothly. Ensure that the
speed measuring encoding disk and speed measuring optocoupler do not rub each other.
7.
When installing the pump door board, ensure that the surface of the display screen and transparent
window of the veneer is clean without dust.
8.
All cables and wires must be connected correctly and firmly. The wire arrangement must be tidy,
and the wires must be bundled and fastened as requested