AFG/AFTI Mk 4
Service and Repair Manual
Number 022-614
Issue Number: 09
Page 32 of 47
- Check added to ensure all calibration data stored correctly
- Message displayed when calibration data stored correctly
- Disable skipping of calibration steps if gauge type changed
2) Memory Counter
- Memory counter stored in eeprom after memory cleared
3) Digmatic
- Timeout reduced
4) AFTI Selection
- AFTI can be selected without gauge having to be set up as AFG
first
Implemented in version 1.01 from serial number
08-0118-03
(Three page menu version)
-------------------------
1) Calibration Routine
- PC Commands working in Calibration Routine. CTRL+A, CTRL+B,
CTRL+C, CTRL+D, CTRL+E received via RS232 enabled in cal routine
to simulate keys. Enabled from load cell size select page up to
(but not including) new 'Store Data' page
- All interrupts disabled in correct place as gauge could 'lock up'
if auto-off/backlight enabled
- Calibration steps can be skipped using the TXD key
- Extra stage at end of calibration routine added to prompt user to
store data
2) X-Constant
- Floating point rounding error fixed when determining decimal
places for new X-constant value
- Maximum number of allowed decimal places reduced
- Loads stored and read with correct units and decimal places when
X-constant enabled/disabled
- Correct data formatting used after X-Constant enabled/changed
- Units key now working correctly within X-Constant page
3) Configuration Status request
- Unrequired blank spaces before Stand Cycles removed
4) Cycle Counter
- Cycle Counter displayed as default if memory counter also used
5) Stand Cycle Limits Menu
- Menu page now exited properly with MAX key