![Mecmesin AFTI Скачать руководство пользователя страница 25](http://html.mh-extra.com/html/mecmesin/afti/afti_service-and-repair-manual_1759020025.webp)
AFG/AFTI Mk 4
Service and Repair Manual
Number 022-614
Issue Number: 09
Page 25 of 47
Futek load cell wiring:
Pin1: +ve Excitation - Green
Pin2: +ve Signal - White
Pin3: -ve Signal - Red
Pin 4: - ve Excitation
– Black
SWJ load cell wiring
Pin1: +ve Excitation - Red
Pin2: +ve Signal - Green
Pin3: -ve Signal - White
Pin 4: - ve Excitation
– Black
16. Load cell stud fitting.
NOTE: The replacement load cell stud and retaining nut type will depend on the
capacity of the load cell fitted:
Gauge
Load cell
Stud
Nut
2.5N
400-175
406-410
N/a
5N
406-410
406-410
N/a
10N
400-006-S01
406-408
420-138
25N
400-007-S01
406-408
420-138
50N
400-008-S01
406-408
420-138
100N
400-024-S01
406-408
420-138
250N
400-107-S01
406-408
420-138
500N
400-016-S01
406-408
420-138
1000N
900-302
406-409
420-316
2500N
900-164-V02
Pinned complete assembly
Pinned complete assembly
63. Degrease the load cell stud and nut and allow to dry.
64. Put a small amount of Loctite 243 on to the stud (opposite end to small hole).
During the load cell assembly routine, handle the load cell by central beam
only. Do not handle by any other means.