AFG/AFTI Mk 4
Service and Repair Manual
Number 022-614
Issue Number: 09
Page 16 of 47
24. Press Units/Menu key again to
display page 2 of menu
25. Press TXD/UP key once until arrow
points to “CALIBRATION”
26. Press Reset/Enter key to enter
Calibration mode
27. Press TXD/UP key three (3) times
until 3000 is displayed