AFG/AFTI Mk 4
Service and Repair Manual
Number 022-614
Issue Number: 09
Page 31 of 47
2) Remove load sign change feature in calibration routine. (Flyspray ID76)
- The calibration routine contains a feature that allows the load sign to be
changed if the load cell has been wired incorrectly. However, although it alters
it for the digital side of the gauge, the polarity of the analogue output remains
incorrect. Therefore the feature needs to be removed so that if a load cell is
wired incorrectly, the load cell should be removed and wired the correct way.
Implemented in version 1.04 (DCR1464) from serial number 08-0168-08
-------------------------
1) Auto-off Eeprom address (Flyspray ID27)
- Auto-off eeprom address moved as overwrote Rate in eeprom
causing gauge to enter 'High Rate' mode which was perceived
as noise in the Orbis
2) Zero Blanking (Flyspray ID28)
- Zero blanking added for high rate so that any load to be
displayed that is less than the accuracy of the gauge is
displayed as zero
3) Loadbar sign (Flyspray ID29)
- Loadbar now always correct appearance
4) Calibration in non-SI units (Flyspray ID40)
- Calibration routine allows calibration in units other than
N and Nm
Implemented in version 1.03 (DCR1444)
-------------------------
1) Calibration Routine
- Calibration routine not entered if Smart connected
2) Turning on with Smart connected
- Delay added to allow hardware to switch to Smart load
3) Torque Smarts
- Smart Torques displayed as correct type and not as force
Implemented in version 1.02 (DCR 1434) (Not Released)
-------------------------
1) Calibration Routine
- Low battery symbol check added to calibration routine