32. Press Reset/Enter key to enter the
third number
33. Press TXD/UP key nine (9) times
until 9 is displayed.
34. Press Reset/Enter key to enter the
forth number
The LCD should then display page 2 of the Calibration menu correctly.
If the text is still inverted
3178
should be entered as above. Restart from instruction
6
35. Press Max/Esc key to return to
‘Home’ page
36. Turn PCB OFF and then back ON.
Text now should still be displayed
correctly
37. DIP Switch SW1 should then be
returned to the ON position.
Procedure ends.