![Mecmesin AFTI Скачать руководство пользователя страница 12](http://html.mh-extra.com/html/mecmesin/afti/afti_service-and-repair-manual_1759020012.webp)
AFG/AFTI Mk 4
Service and Repair Manual
Number 022-614
Issue Number: 09
Page 12 of 47
9. Sounder replacement (328-005)
12. Remove PCB as per instructions 1
to 11 in section 6.
13. De-solder the sounder from the
PCB and replace, observing the
correct polarity. The positive
(longer) pin should be inserted
through the hole with a square
solder pad. After soldering, trim off
the excess sounder pins
14. Remove the sticker from the
sounder
15. Test the Sounder. Turn the instrument on. The audible alarm should sound.
Reassembly is the reverse of the above procedure