![Mecmesin AFTI Скачать руководство пользователя страница 24](http://html.mh-extra.com/html/mecmesin/afti/afti_service-and-repair-manual_1759020024.webp)
AFG/AFTI Mk 4
Service and Repair Manual
Number 022-614
Issue Number: 09
Page 24 of 47
15. Load cell removal
58. Remove front panel assembly as
per instructions 1 to 7 in section 6.
59. BEFORE REMOVING LOADCELL
make careful note of the way it is
fitted
– the routing of its wires,
positioning and use of sticky pads
etc
– The replacement load cell
MUST be refitted in an identical
manner.
60. Load cells for 2.5N
– 500N gauges
are retained with 2 off 6/32 UNC x
5/8” socket cap screws (420-265).
Remove with appropriate Allen
key. 2.5 & 5N load cells have
spacers (408-392) fitted.
61. Load cells for 1000N and 2500N
gauges are retained with 2 off M8 x
23 hex head screws (421-401).
Remove using appropriate socket.
62. The 2500N load cell maybe found
to have been pinned into position.
If so, a replacement 900-164-V02
AFG MK4 2500 PINNED L/C
ASSY is required.