6 Useful Features
27
R4G/R4GE series
6
Useful Features
6-1
Master/Slave Setting (R4G series with the -LGmb option)
Set the role of "Master" or "Slave" with the [Master/Slave] selector switch on the rear panel.
It is set to "Master" at factory shipment.
When the [Master/Slave] selector switch is set to "Slave", only the remote
operation is enabled and the voltage/current setting dial operation on the
front panel is
disabled
.
As the master unit sent various settings to its slave units when it is turned
on, make sure to turn on the slave units first in the case of the
master/slave unit configuration.
6-1-1
Description of Master/Slave Function
Set the [Master/Slave] selector switch of a unit selected as the master to "Master" and the switch of a
unit selected as the slave units to "Slave", and connect between the CO-M cable connector (OUT) of
the master unit and the CO-M cable connector (IN) of the slave unit using the supplied CO-M cable.
(For details, refer to [6-1-2 Master/Slave Connection Diagram].)
Slave units immediately follow and vary each operation of the master unit such as the OUTPUT
ON/OFF (when the [OUTPUT ON/OFF] switch is set to ON of the slave units), output voltage setting,
output current setting, OVP, OCP.
Use all "+ Sense" and "- Sense" terminals when using the remote sensing function.
Precautions on the Master/Slave Function:
When the master unit detects that a protection function has been activated, the outputs of
all
slave units
connected to the master unit will also be turned off. If a slave unit detected it, only
the output of
the detected slave unit
will be turned off.
In order to turn on the output of a slave unit whose output has been turned off, turn off the
[OUTPUT ON/OFF] switch on the master unit to reset everything, and then turn it back on.
*
The above operation is applied to OVP, -LD (Interlock), and OCP.
In the master/slave function, the master unit sends setting signals to the slave units when it is
turned on. Therefore,
turn on the slave units first to start up the slave units
, and then
turn on the master unit approx. 10 seconds later to start up the master unit. Pay attention to it
when configuring your system. When outputting, press the [SET] switch of the slave units to
check that the preset values of the master unit are set in the slave units before turning the output
ON. (* with delay trigger function in slave remote setting)
The Master/Slave function in this section is controlled by digital signals.
。
Note that if analog
signals are connected only to the master unit, only the master unit will be controlled. (For details,
refer to [7-3 Remote Programming].)
When using analog signals to control all units, connect analog signals to the master unit and
each slave unit respectively. When using analog signals to operate the Master/Slave
configuration, refer to [7-5 Single Control Parallel Operation].
When turning the output on and off with the -LS (remote switch), use the "OUTPUT" signal sent
from the master unit to the slave units.
Do not form a master/slave configuration using different models.
Only the same model can form a master/slave configuration. Mixing different models
for it may cause failure or malfunction
Master
Slave