1 Introduction
3
R4G/R4GE series
1-5
Specifications
Item
Description
Input voltage
120Vac±10%, 50/60Hz, single phase
Output control
(voltage &
current)
R4G series:
Local: Rotary encoder on the front panel
Remote: External control voltage: 0 to 10Vdc or external resistance: 0 to 10kΩ
R4GE series:
Rotary encoder on the front panel
Stability
0.05% of max. output voltage/8H
Temp. coefficient 0.01%/°C (Typ.) of max. output voltage, 0.02%/°C (Typ.) of max. output current
Output display
Output voltage: 4-digit digital indicator, accuracy ±0.3% rdg. + 2-dgt.
Output current: 4-digit digital indicator, accuracy ±0.5% rdg. + 3 dgt.
Protection circuit
Overvoltage protection (OVP), Overcurrent protection (OCP),
Power failure protection (PFP)
Other functions
Common to R4G/R4GE series: Key lock to prevent accidental operation, Last setting
memory
R4G series: ON/OFF by Remote switch, Remote sensing, Status signal output, Delay
trigger function, Multi-set function
Transient
response time
Recovery time: 50μs (at 10%
⇔
100% load regulation)
Operating
temperature
0
℃
to +40
℃
Storage
temperature
-20
℃
to +70
℃
Storage humidity 20% to 80% (no condensation)
Withstand voltage
Between input and output terminals and between input terminals and chassis:
AC1000V/1 min.
Voltage to ground
Models with the output voltage of 80 V or lower: ±250 V-DC
Models with the output voltage of 120 to 360 V: ±500 V-DC
Models with the output voltage of 650V: ±650V-DC
(
“+” or “-“ terminal grounding possible)
Содержание R4G Series
Страница 1: ...F RA 001 2R3 Instruction manual MODEL R4G R4GE series B N 217 9 138 217 9 138 Rev 0 8 ...
Страница 17: ...9 R4G R4GE series 2 6 External Dimensions Unit mm 2 6 1 Width 84 mm Model R4G E6 2 R4G E18 2 R4G E36 1 ...
Страница 19: ...11 R4G R4GE series 2 6 3 Width 210 mm Model R4G E18 10 R4G E36 5 R4G E45 4 ...
Страница 60: ......