4 Normal Operation
22
R4G/R4GE series
LOCK
OCP
SET
OVP/
CV
CC
V
A
OUTPUT
REMOTE
1. In the case of a single operation or an external relay contact can be prepared for each unit:
2. In the case of reversing the above logic:
In the R4G series, the -LS (Remote Switch) logical inversion function is set to on (inverted) in the
startup menu at factory shipment.
When using the -LS (Remote Switch), -LS (Remote Switch) can be used with normal logic by setting
"LS" (Remote Switch) to "oFF" in the startup menu. Use it between PIN 10 and PIN 21 of the supplied
MIL 26PIN male connector.
The R4G series comes with a remote connector, but it is not short-circuited between PIN 1 and
PIN 21.
Internal circuit
+S
(common)
TB1
10
21
TTL signal
4.7kΩ
100 Ω
+S
(common)
-
+
External connection
External relay
TTL LOW
=
OUTPUT: ON
TTL HIGH
=
OUTPUT: OFF
or
External relay CLOSE = OUTPUT : ON
External relay OPEN = OUTPUT : OFF
Turn the current setting dial to select "on" or
"oFF" of the -LS (Remote Switch) function.
VOLTAGE
CURRENT
Current setting dial