5 Basic Operation
26
R4G/R4GE series
5-3
Parallel Operation
The output current can be increased by connecting two or more of the same units in parallel.
.
When using the unit at constant voltage:
Make sure to set the output voltage of each unit to the same.
Initially, E1 with a high output voltage supplies the current, and
the mode will shift to the constant current range as the current
increases, and the output voltage decreases.
Then, E2 with the second-highest output voltage supplies the
current.
Consequently, the difference in output voltage, such as E1>E2,
becomes a fluctuation for the load.
When using the unit at constant current:
Set each unit to a constant current state.
The output current will be the sum of the output currents of each unit.
Do not apply a higher voltage than the rated one to the output of each unit.
A few mA of current will flow back to the units with lower output voltage.
Parallel Operation
Load
E
2
I
2
I
1
E
1
FINE
REGULATED DC POWER SUPPLY
POWER
LOCK
OCP
SET
OVP/
CV
CC
V
A
OUTPUT
VOLTAGE
CURRENT
R4G18-2
-
+
GND
!
|
○
REMOTE
FINE
REGULATED DC POWER SUPPLY
POWER
LOCK
OCP
SET
OVP/
CV
CC
V
A
OUTPUT
VOLTAGE
CURRENT
R4G18-2
-
+
GND
!
|
○
REMOTE
Содержание R4G Series
Страница 1: ...F RA 001 2R3 Instruction manual MODEL R4G R4GE series B N 217 9 138 217 9 138 Rev 0 8 ...
Страница 17: ...9 R4G R4GE series 2 6 External Dimensions Unit mm 2 6 1 Width 84 mm Model R4G E6 2 R4G E18 2 R4G E36 1 ...
Страница 19: ...11 R4G R4GE series 2 6 3 Width 210 mm Model R4G E18 10 R4G E36 5 R4G E45 4 ...
Страница 60: ......