Turn on the carrier gas, compressed gas and
instrument power.
Restart the software.
In the leak test dialog box select SV4 and HV
Leak test around the brass connector with a helium
leak detector.
If a leak is detected, gently tighten the brass
connector 1/8 turn with a 7/16” spanner and check
for leaks again. Repeat until no leak is detected.
DO NOT OVERTIGHTEN OR YOU RISK
BREAKING THE COLD TRAP AGAIN.
carrier gas, compressed gas and
In the leak test dialog box select SV4 and HV
Leak test around the brass connector with a helium
ently tighten the brass
n with a 7/16” spanner and check
for leaks again. Repeat until no leak is detected.
DO NOT OVERTIGHTEN OR YOU RISK
BREAKING THE COLD TRAP AGAIN.