![Markes International Unity 2 Скачать руководство пользователя страница 59](http://html.mh-extra.com/html/markes-international/unity-2/unity-2_user-manual_1727982059.webp)
QUI-1057
Page 59
Markes International Ltd
T:
+44 (0)1443 230935
F:
+44 (0)1443 231531
E:
www.markes.com
You will feel increased resistance as
the cold trap pushes into the seal at
the valve end.
Once the trap is properly inserted the
glass flange should be approximately
1 mm from the black plate
Keep a firm hold of the trap
pneumatics assembly during these
steps – if it is allowed to twist it can
snap the cold trap.
Bring the pneumatic assembly back
and relocate the screw in the slot.
Push the assembly gently back in the
horizontal plane guided by the screw.
THE CONNECTOR CAN SNAP
THE END OF THE QUARTZ COLD
TRAP
Apply gentle steady pressure to push
the trap into the sealing O-ring located
inside the steel trap connector.
Retighten the screw firmly