5.3.3
Removing / replacing a broken cold trap
The carrier gas supply and compressed air to
the instrument and the instrument itself must
be turned off before starting.
Remove the visible end of the cold trap using the
method described in Section 3.2
Unscrew the 2 screws securing the front panel.
Remove the front panel.
Slide the side panels towards the front of
instrument then lift away to remove.
Remove the air lines that supply the heated valve by
pulling the white tube while pressing the black lip of
the connector.
Remove the split tube and any sample tubes /
interconnects.
Removing / replacing a broken cold trap
The carrier gas supply and compressed air to
instrument itself must
Remove the visible end of the cold trap using the
Unscrew the 2 screws securing the front panel.
Slide the side panels towards the front of the
Remove the air lines that supply the heated valve by
pulling the white tube while pressing the black lip of
Remove the split tube and any sample tubes /
1.
2.
QUI-1057
Page 61
3.