![Markes International Unity 2 Скачать руководство пользователя страница 53](http://html.mh-extra.com/html/markes-international/unity-2/unity-2_user-manual_1727982053.webp)
QUI-1057
Page 53
Markes International Ltd
T:
+44 (0)1443 230935
F:
+44 (0)1443 231531
E:
www.markes.com
If they are required to be changed they should
be hooked out with the O-Ring extraction tool
(SERZ-0351) from the shipping kit.
Use the same pointed implement to hook out
the PTFE filter disks.
The clean filters (U-DISK1) should be pushed
in with the O-Ring insertion tool (SERZ-0285)
or with any clean, flat-ended metal rod of
suitable diameter.
New O-rings should also be pushed into
position using the O-Ring insertion tool
supplied in the shipping kit.
Also use the O-Ring insertion tool to smooth
around the inner diameter of the O-Ring as it
is being pushed into place to avoid distortion.
5.2.6
Changing the charcoal split tube filters
If you are not recollecting the split effluent for subsequent analysis, the charcoal filters will trap the split
effluent to prevent it being vented to the atmosphere. Therefore the filters will require changing at routine
intervals.
SERUTD-5065 Split Filter Tube Packed with Charcoal (UNITY 2)
SERAAA-1600 Split Filter Tube, Glass, Packed with Charcoal (Air Server systems)
It is suggested that the filters are repacked with conditioned charcoal every three months. Charcoal filters
are loaded in the same manner as tubes, with the grooved end (or fritted end of the glass tube) facing the
back of UNITY.
Once the split tube has been isolated from the carrier gas flow by selecting the ‘Change Split Tube’ icon,
remove the charcoal tube by lifting the black knob of the sealing mechanism and using the Tube Extractor
tool (SERUTD-5062) to grip the charcoal tube and remove it from the seal.
5.2.7
Changing DiffLok cap o-rings (UltrA)
It is rare for the O-rings in the DiffLok caps (U-COV10 / U-COV100) to require replacing, however if
necessary (O-rings have become worn or badly contaminated), then it is a simple operation to remove the
O-rings and replace using the following O-ring extraction tool (SERZ-0351) and insertion tool (SERZ-0285).
Spare DiffLok caps are available as follows:
C-DLS10 Cap, DiffLok, inert coated, pk 10 - for use on hot end