2 Safety
| HCU 40 |
Instructions for Use | 1.0 | EN | 01
Copyright Maquet Cardiopulmonary GmbH
2.2.3 Handling the HCU 40
WARNING!
n
During an application, only use devices and equipment which are functioning
perfectly.
n
Do not connect equipment which does not form part of this system.
n
Do not touch the touchscreen with sharp or pointed objects.
n
Only use the approved liquids as per the Instructions for Use in and on the
HCU 40.
n
Do not touch the plugs of the HCU 40 as electrostatic charges and moisture
may cause damage.
n
Do not disconnect any plugs or cables from the HCU 40 or the control unit
(CU) during operation.
n
If the CAN connection cable between HCU 40 and control unit (CU) is not
connected during operation, immediately reconnect the cable (⇨ "Connecting
a Control Unit", page 39).
n
If a cable proves to be defective, replace it with a cable which functions
correctly.
n
If a plug is faulty, do not operate the device.
n
If the touch screen of the control unit (CU) does not display anything or fails
to react, the CU must be reset. Disconnect and reconnect CAN connection
cable between HCU 40 and CU. Check the set values.
n
If, during a power failure, no visual or acoustic alarms are emitted by the
HCU 40, it is possible that the independent power supply (UPS) is defective.
Have the device checked/repaired by authorized service personnel. Use a
replacement device in an emergency.
2.2.4 Water Circuits and Heat Exchangers
WARNING!
n
Switch the device off before intra-hospital transportation. Remove all
connected cables.
n
Observe the shipping information (⇨ "Send Device to Authorized Service
Point", page 131).