| HCU 40 |
4 Operation
| 47 |
Instructions for Use | 1.0 | EN | 01
Copyright Maquet Cardiopulmonary GmbH
▶
The HCU 40 stops the activated pump and displays the
[Pump not running]
symbol.
4.2.4 Using Hotkeys
The following instructions apply for the patient water circuit and the cardioplegia
water circuit. Hotkey
[Warm patient water circuit]
is shown.
1
Touch a hotkey in order to use its values, e.g.,
[Warm patient water circuit]
.
▶
The touchscreen displays the values
set for the hotkey: T
set
" for setpoint
temperature,
[ΔT]
for setpoint
gradient.
▶
The touchscreen displays a warning symbol if the setting may be harmful
to health (see table below).
2
Ensure that the selected setting is suitable and safe for the patient and the
current situation.
3
To confirm and use the displayed values, touch the
[Confirm]
symbol.
▶
In this case, the HCU 40 adopts the values of the hotkey as the setpoint
temperature and setpoint gradient.
▶
The setpoint gradient is only used if external temperature sensors are
connected. Gradient mode is deactivated with the value
[Off]
.
Or
4
If you do not want to use the values, touch the
[Reject]
symbol instead.
Cause
Message
Setpoint temperature
> 39.4°C
Temperature setting may be harmful to health!
Setpoint temperature
< 33.0°C
Gradient > 8.0°C