7 Cleaning, descaling and disinfection
| 121 |
Instructions for Use | 1.0 | EN | 01
Copyright Maquet Cardiopulmonary GmbH
25 Touch the
[Next]
symbol.
26 Close the water pump stopcocks, reattach the yellow caps and disconnect the
drainage tube.
Or
Disconnect the drainage tube.
27 Touch the
[Next]
symbol.
Rinsing cycle 1
1
Disconnect the cleaning connectors and disinfect the Hansen couplings and
the cleaning connectors.
2
Connect the cleaning connectors.
3
Fill the HCU 40 up to the white mark in the tank with sterile filtered water (1
cm above the cooling plates).
4
Touch the
[Next]
symbol.
▶
The HCU 40 de-airs the cardioplegia and patient water circuits.
5
Disconnect the cleaning connectors.
6
Place the open ends of the tubes (without touching any contaminated
surfaces) in an outflow and open the stopcocks of the patient and cardioplegia
water circuits.
7
Touch the
[Next]
symbol.
▶
The HCU 40 pumps the water out.