Message
Possible conse
quences
Possible cause
Possible action
Pat.: Water flow low.
If error is not rectified:
circulation is switched
off, and therefore tem
perature regulation not
possible.
Outlet valve closed.
Open outlet valve.
Tube kinked.
Remove kink.
Blanket operated with
out pressure reducer.
Connect pressure re
ducer. Use replace
ment device.
Pat.: No ice detected!
After 6 hours of com
pressor running time,
no ice formation de
tected in the patient
tank. Device defective!
Use crushed ice.
No ice formation, and
thus no cooling possi
ble.
Ambient temperature or
tank temperature too
high.
Check ice formation
and restart if neces
sary. Place device in
cooler environment. In
case of emergency,
use crushed ice. Use
replacement device.
Pat.: External temp.
sensor defective!
Gradient mode not
available.
External temperature
sensor defective!
Change external tem
perature sensor.
Control unit defective.
If the error occurs
again: Notify authorized
service. Use replace
ment device.
Pat.: Ext. temp. sensor
disconnected!
Gradient mode not
available.
External temperature
sensor disconnected.
Reconnect sensor or
confirm disconnection,
if necessary.
Pat.: Ext. temp. out of
range!
Gradient mode not
available.
External temperature
sensor defective!
Check the actual tem
perature. Change ex
ternal temperature
sensor.
Control unit defective.
If the error occurs
again: Notify authorized
service.
Pat.: Tank temp.
sensor defective!
Check ice visually.
Displayed tank temper
ature is incorrect. There
may not be any cooling
of the patient.
Tank temperature
sensor defective!
Notify authorized ser
vice. Use replacement
device.
Mains voltage too high. Device can overheat
and be damaged.
Supply voltage too high
or voltage measure
ment incorrect.
Check supply voltage
at HCU 40 (CU). In
case of doubt, contact
service.