3 System Description
| HCU 40 |
Instructions for Use | 1.0 | EN | 01
Copyright Maquet Cardiopulmonary GmbH
Left side: Patient water circuit
Right side: Cardioplegia water circuit
Function
Meaning
Deairing.
"Deairing the Circuits", page 56)
Emptying the tubes.
Empty the tubes of the patient and cardioplegia water cir
"Emptying Water Circuits", page 58)
Compressor control.
"Compressor Control", page 31)
3.8.1 De-airing the Circuits
WARNING!
With the "De-airing" function, ice-cold water is pumped from the tank into the
tubes and the heat exchanger. The perfusion circuit in which the heat exchanger
being de-aired is integrated must not be operated, as it could provoke a cardiac
arrest if a patient is connected.
This function can be used to de-air the patient circuit and cardioplegia water
circuit independently of each other prior to the application. To this end, the
touchscreen displays the symbols for de-airing in the "Functions" screen.
Function
Meaning
Deairing can be started.
To start the deairing, touch the symbol of the circuit that
you want to deair.
Deairing running.
To stop the deairing, touch the symbol.
Deairing completed, tem
perature regulation running.
The circuit is deaired; the temperature of the water is be
ing regulated.
Deairing and temperature
regulation completed.
The outlet temperatures have attained the setpoint tem
peratures.