87
MAGNAMED TECNOLOGIA MÉDICA S/A
Adjustable Parameter Specifications
Table 18
– Adjustable Parameters
Parameter
Specification
Resolution
Unit
Tidal Volume
10 to 3000
Pediatric
10 to 100: 5
mL
100 to 300: 10
Adult
100 to 1000: 10
1000 to 3000: 50
Breathing Rate
(1)(2)
0 to 200
Neonatal
0 to 200:1
rpm
Pediatric
0 to 200:1
Adult
0 to 100:1
Rise Time
0 to 2,0
0.1
s
Pause
0 to 70
10
%
0 to 2,0
0,1
s
Inspiratory and Limit Pressure
0 to 120
1
cmH
2
O
∆PS
0 to 120
1 cmH
2
O
cmH
2
O
PEEP
0 to 50
1 cmH
2
O
cmH
2
O
Pressure trigger
0,0 to -20
0.00 to -2.0 : - 0,2
cmH
2
O
-0.05 to -20 : - 1
Flow trigger
0,0 to 30
0.00 to 1.0 : 0,1
L/min
1,0 to 30,0 : 0,5
Cycled-Flow
5 to 80
(max. 3 s)
5
%
O
2
Concentration
21 to 100
1
% vol
Inspiratory Time
0,05 to 30
0.05 to 0,70 : 0,01
s
0.70 to 1,00 : 0,05
1,0 to 30,0 : 0,1
Inspiratory Flow’s Waveform
Square,
Descending or
Decelerated,
Ascending or
Accelerated,
Sinusoidal or sinusoid
---
---
CPAP
0 to 50
1
cmH
2
O
High Pressure (APRV/DualPAP)
5 to 90
1
cmH
2
O
Low Pressure (APRV/DualPAP)
0 to 45
1
cmH
2
O
High Time (APRV/DualPAP)
0,10 to 59,9
0,10 to 0,70: 0,01
s
0,70 to 1,00: 0,05