CPAP/PS
–
Continuous Positive Airway Pressure/ Pressure Support
1 and 2 Represents spontaneous cycles with pressure support on ZERO.
3, 4, and 5 -
Represents the patient’s spontaneous respiration cycles with a pressure support different
from zero. The pressure support’s T
RISE TIME
can be adjusted in order to mitigate the initial flow. The cycling
flow may be adjusted to a value between a peak flow of 25% to 75%.
6. In the event the patient enters into apnea, the ventilator will present this condition through an alarm
on the screen’s message and alarms area after T
APNEA
and will initiate selected back-up ventilation, according
to the programmed configurations and parameters.
•
The set supporting pressure (ΔPS) is a value above PEEP. Therefore, the supporting inspirato
ry
pressure will be the sum of PEEP and ΔPS.
•
Default values are only an initial reference.
•
Readjust the ventilation parameters according to the patient’s needs