50
MAGNAMED TECNOLOGIA MÉDICA S/A
9
Preventive Maintenance
•
The symbol
displayed on the screen ventilator, it indicates that the equipment entered the
preventive maintenance period. For the equipment ICU this period is 5,000 hours or 12 months,
whichever occurs first.
•
Schedule preventive maintenance only by the technical service authorized MAGNAMED.
•
Before sending the equipment to the technical service note STRICTLY the process of cleaning
and disinfection.
Verifications
The following verifications must be carried out on a daily basis and whenever the equipment is to be
used:
A. Cleaning equipment;
B. Integrity of power source AC/DC converter;
C. Operation of alarm systems, including audio;
D. Filters Air/O
2
installed and unobstructed;
E. LCD;
F. Loaded Batteries;
G. Touchscreen;
H. Turn and confirm button;
I.
Correct respiratory circuit installation (including the exhalation valve diaphragm);
•
A daily check should be carried out with the disconnected ventilator of the patient.