44
MAGNAMED TECNOLOGIA MÉDICA S/A
Problem
Possible Causes
Solutions
Incorrect PEEP control
Exhalation valve calibration.
Restart the equipment and calibrate the
exhalation valve
Curves and trend values do
not appear or are incorrects
Failure in the clock system setting.
Request for Technical Assistance
Equipment does not start
ventilation
Equipment in STAND BY
Press the STAND BY button during 1
second to start the ventilation
Test failure of proportional
valves.
Pressure of the gas network below
the lower limit.
Check and reset the network pressure
to reach the specified range.
It is not possible to active the
automatic adjust of the alarms
Equipment in STAND BY mode
The automatic adjust of the alarms may
only be calculated if the ventilator is
cycling.
Press STAND BY button during 1
second to start the ventilation and wait
the ventilation stabilization to activate
the automatic adjust.
Test failure of distal flow
sensor
The output of the breathing circuit
was not occluded.
Restart the equipment and redo the self
test with the respiratory circuit closed
Test failure of the exhalation
valve
The output of the breathing circuit
was not occluded.
Restart the ventilator and redo the self
test with the respiratory circuit closed.
The diaphragm of the exhalation
valve is misplaced.
Reposition
the
exhalation
valve
diaphragm, restart the machine and
redo the self-test.
The parameter setting returns
to the previous value.
The adjust was not confirmed
Confirm the adjust by pressing the
button or by touching in the parameter
Audible alarm inoperative
Mute alarm is active
Turn off the mute alarm
Electronic failure
Request for Technical Assistance
Backup is not active
Apnea time alarm is off
Set an interval for apnea alarm
•
Never use the equipment if a problem cannot be solved.